Masimo SafetyNet Alert Manual Del Usuario página 53

Ocultar thumbs Ver también para SafetyNet Alert:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 305
Symptom
Potential Causes
Delayed data
updates
User alarms do not
appear during
events
Masimo SafetyNet
Alert App does not
detect that the
sensor is applied to
the hand
Wi-Fi is not correctly
configured.
Incorrect user currently
logged into Masimo
SafetyNet Alert App.
Wi-Fi network not
available.
System settings have
changed.
Sensor not properly
placed on user.
Sensor not properly
paired to Bedside
Station.
Damaged sensor.
Internal failure.
Next Steps
Check that the wireless feature is correctly
configured. Refer to the smart phone's
Operator's Manual or Directions For Use.
Check network settings and availability. You may
need to call network provider for further
assistance.
Restart Masimo SafetyNet Alert App and login to
the system. See
Contact Masimo Customer Support. See
Customer Support
Ensure Masimo SafetyNet Alert is connected to
Bedside Station Setup
Wi-Fi. See
Ensure the correct user is logged into Masimo
SafetyNet Alert App.
Restart Masimo SafetyNet Alert and login to the
system.
Ensure alert notification settings are turned on.
Contact Masimo Customer Support. See
Customer Support
Reapply the sensor.
Pair the sensor with Bedside Station. See
Pairing the Masimo Sensor
Replace sensor.
Turn Bedside Station Off and On by unplugging
the device then plug it in to start up
Contact Masimo Customer Support. See
Customer Support
51
on page 69.
on page 21.
on page 69.
on page 26.
on page 69.
Tabla de contenido
loading

Tabla de contenido