General Description

The SSCOR Quickdraw is a non-sterile hand held 12V DC battery powered portable suction device to be used by
professional personnel trained in Emergency Care techniques of constant suctioning to clear the airway by removing
bodily fluids and particulate matter. Suction power can be regulated when full power may be considered harmful to the
patient. The SSCOR Quickdraw is transportable in a road ambulance, or in a fixed or rotary winged aircraft.
A fully charged alkaline battery will power the unit continuously for 180 minutes (±10%). After running for 180 minutes
(±10%) the unit will run at reduced power and the red battery condition LED will light indicating it is time to replace the
battery. When the red LED begins to blink, only a few minutes run time remain. Install a new battery in the Quickdraw.
The battery is a single use battery and is intended to be discarded once it has been discharged. Dispose of the battery
according to local / regional / national requirements for the disposal of electronic waste. Battery condition is automatically
monitored and indicated by an indicator light on the side of the chassis.
The 80615 10xAAA Battery Holder with fresh alkaline batteries at full capacity and voltage will power the Model 2403
Series Quickdraw suction device continuously for approximately 60 to 100 minutes depending on the level of battery
capacity, battery voltage, and the brand of battery. As the battery capacity begins to decline, the red battery condition LED
will light indicating it is time to replace the batteries in the battery holder. When the red LED begins to blink, immediately
install fresh batteries into the battery holder

Battery Warning

The 80611 sealed lead acid battery can be used in the alkaline version of the Quickdraw (Model 2403 Series). The LED
will properly display the battery condition. The 80611 sealed lead acid battery will power the alkaline version of the
Quickdraw (Model 2403 Series) for 30 minutes (±10%), but cannot be charged within the unit.

Warranty

SSCOR warrants that each new product is free from defects in material and workmanship under normal use and service
for a period of one year from date of purchase. This warranty gives you specific legal rights and you may also have other
rights that vary from jurisdiction to jurisdiction. For countries where minimum warranty terms are determined by statute,
the warranty term is the longer of the statutory period or the term listed above. Batteries, disposable items including
collection canisters, patient tubing and catheters are excluded from this warranty. See the SSCOR Warranty for terms and