batt
...
appears in the display?
– The rechargeable battery is so severely discharged that weighing is no longer possible.
Recharge the battery.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display shows
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, interrupt the power supply by briefly removing the rechargeable
battery. If this measure is equally unsuccessful, inform the Service department.
... the weighing range won't switch?
– Take the load off the scale if necessary.
– Ensure that no special function is active.
9. Maintenance
Have recalibration performed in accordance with national regulations. Servicing and recali-
bration may be performed solely by authorised staff. Contact your service agent or seca
Customer Service.