Ocultar thumbs Ver también para LOCATOR:

Enlaces rápidos

Instructions for Use: LOCATOR
Male Removal Tool
NOTE: The Abutment Holder Sleeve for use with the LOCATOR
from the Core Tool in a 4 pack or 20 pack, (Zest order numbers 8394 and 8394-20). To use the driver with sleeve start
with Step 1. To remove and replace the LOCATOR
STERILIZATION
LOCATOR
Core Tools (in the disassembled state only) may be sterilized by Autoclave or Dry Heat sterilization using
®
the following parameters:
1. Autoclave sterilize using 121° C (250° F), (15-20 psig at sea level), for 40 minutes minimum.
2. Dry Heat sterilize using 170° C (338° F) for 2 hours minimum.
NOTE: The Abutment Holder Sleeve has been validated to withstand five (5) autoclave sterilization cycles using
parameter no.1 above. The Sleeve may also be sterilized using a liquid chemical sterilant following manufacturer's
instructions. The liquid chemical sterilant must be approved for Sterilization not just High-Level disinfection and
it must be compatible with polysulfone material.
1. The Sleeve (Zest order number 8394) slips onto the driver end of LOCATOR
LOCATOR
Abutment onto the driver, allowing the driver/sleeve and abutment to be held vertically without abutment
®
falling off the driver, making it easier to deliver the abutment to the patient's implant.
2. Loosen the (8397) Male Removal Tool a full 3 turns counter clockwise (you will see a visible gap).
3. To remove a LOCATOR
and push straight in to the bottom of the nylon male. Then tilt the tool so that the sharp edge of the tip will grab hold of
the male and pull it out of the cap.
4. To discard the nylon male from the tip on the Core Tool; point the tool down and away from you and tighten the Male
Removal Tool clockwise back onto the Core Tool. This will activate the removal pin and dislodge the nylon male from
the tip end of the Male Removal Tool.
5. Separate the Male Removal Tool section from the LOCATOR
remaining two sections to place a new nylon male into the empty titanium metal housing.
L8009-TM REV. D
CORE TOOL (with Abutment Holder Sleeve)
®
Gap
nylon male from the titanium metal housing; simply insert the tip into the cap/male assembly
®
Male Seating Tool
Core Tool is pictured above. It is available separately
®
Male start with Step 2.
®
Core Tool and use the Male Seating Tool end of the
®
Male Seating Tool End
Abutment Driver Sleeve
Core-Tool, and is designed to hold a
®
LOCATOR
is a registered trademark of ZEST IP Holdings, LLC
®
loading

Resumen de contenidos para Zest LOCATOR

  • Página 1 Core Tool is pictured above. It is available separately ® from the Core Tool in a 4 pack or 20 pack, (Zest order numbers 8394 and 8394-20). To use the driver with sleeve start with Step 1. To remove and replace the LOCATOR Male start with Step 2.
  • Página 2 4 ó 20 unidades (números de pedido Zest 8394 y Utensile l in una confezione da 4 o da 20 (Numeri di ordinazione Zest 8394 e commander indépendamment de l’instrument de pose par paquet de 4 ou 8394-20).