Error
Messages
The measurement
results are unreliable
due to motion of the
patient during mea-
Readout
surement. The stated
accuracy of measure-
and "A"
ment is invalid.
alternately
flashing
The pulse rate exceeds
the upper limit (230 bpm)
of the measurement
Flashing
range.
The pulse rate is below
the lower limit (30 bpm)
of the measurement
Flashing
range.
The oxygen saturation
(SpO
lower limit (70%) of the
guaranteed accuracy
Oxygen
range.
saturation
The stated accuracy
readout
of oxygen saturation
flashing
measurements is
invalid.
Cause
) is below the
2
Solution
Properly insert a fingertip
1
into the finger holder. Avoid
moving the finger in question.
Warm the fingertip and/or
2
shut out surrounding light or
electromagnetic noise.
Measurement cannot be per-
formed when the upper limit
of the measurement range is
exceeded.
Measurement cannot be per-
fomed when the lower limit of
the measurement range is not
reached.
This message indicates that the
oxygen saturation is out of the
range in which the stated ac-
curacy of measurements is valid.
E 5