Changing Channels In The Ad-Hoc Mode; Changing Channels On The Milling Unit - Sirona CEREC MC XL Manual Del Usuario

Ocultar thumbs Ver también para CEREC MC XL:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Sirona Dental Systems GmbH
Operating MC XL via WLAN in the Ad-hoc mode CEREC / inLab MC XL
Opening the Internet Explorer
Opening the Internet Explorer
IP address of the milling unit WLAN module
IP address of the milling unit WLAN
module
Login of the milling unit WLAN module
Login of the milling unit WLAN
module
61 90 545 D 3439
D 3439.031.03.01.02 06.2007
4.4

Changing channels in the Ad-hoc mode

Channel change must be done separately on each unit in the Ad-hoc mode.
If multiple devices are connected to the system, each unit must be set to the
same WLAN channel.
4.4.1

Changing channels on the milling unit

Click "Start" / "Programs" / "Internet Explorer".
The Internet Explorer window opens.
The IP address of the milling unit WLAN module is the sum of the milling unit
IP address plus 100.
The IP address of the milling unit appears on the display after you switch the
milling unit on by pressing the "Config" button.
Example:
IP address of milling unit
IP address of WLAN module
1.
Enter the IP address of the milling unit WLAN module in the address line
of the Internet Explorer window.
The login window of the milling unit WLAN module then opens.
4 Final work
Changing channels in the Ad-hoc mode
192.168.230
.120
+100
192.168.230
.220
29