Tabla de contenido
Idiomas disponibles

Idiomas disponibles

NO PULSE
• No battery installed - Insert batteries.
• Wrong battery polarity - Follow the instructions on page 5.
• Low battery - Replace with new batteries.
LED NOT FUNCTIONING CORRECTLY
• Circuit not working properly - Remove battery for
1 minute. Then replace battery and try unit again.
• Low battery - Replace with new batteries.
• Keep the transparent plastic pad backings and always
replace after use.
• Avoid touching the gel side of the pads as this will reduce
the adhesiveness of the surface.
• You can prolong the life of the adhesive surface of the pads
by applying a small amount of water to the gel.
TR OU B LE S HO OTI NG
WEAK PULSE
• Low intensity - Increase by pressing the ON/UP button.
• Low battery - Replace with new batteries
• Aged gel pads - Replace the gel pads.
RED SKIN AREA AFTER USE
• Over use - Reduce duration and/or frequency of use.
• Damaged or dried pad - Dampen the gel with a very small
amount of water. If this does not work replace gel pads.
LOOSE ATTACHMENT OF THE PADS
• The pads are worn out - Replace pads.

U N IT M AIN TE NA NCE

• Gel pads last for approximately seven uses.
• If the unit is left unused for a long period of time, remove
the batteries to prolong their life.
• Do NOT dissemble or repair. There are no user serviceable
parts inside the unit. If you have any problems using the
product, contact our Customer Care Department.