VP472
module to connect pyranometers or albedometers. The measurements
produced during the time by a pyranometer or an albedometer can be
taken, verified and stored. The signal produced by the thermopile can be
read in mV or in W/m
, the net radiation of the albedometer is read in W/m
2
The thermopile sensitivity can be set from 5000 to 30000nV/(Wm
between 5 and 30μV/(Wm
-2
VP473
module for reading the continuous voltage. When connected to the output
of a transmitter with voltage signal it can read and take the relevant value.
Measuring range: ±20Vdc. Input impedance: 1MΩ.
IP472
module for mA reading of continuous current. When connected to the output
of a trasmitter with current signal, it can read and take the relevant value.
Measuring range: 0...24mA. Input impedance: 25Ω.
PP471
module for measuring absolute, relative and differential pressure. It can be
connected with pressure probes TP704 and TP705 series. It measure the
instantaneous value and peak value of pressure. The module is complete
with 2m cable and DIN 45326 8 pole female connector.
Accuracy: ±0.05% of full scale
Peak accuracy: ±0.5% f.s.
Probes complete with SICRAM module
Pt100 sensor temperature probes
TP472I wire Pt100 immersion probe. Tube Ø 3 mm, length 300 mm. 4 wire cable 2m long.
Working range: -196°C...+500°C.
Accuracy: ±0.2°C (-196°C...+350°C) / ±0.4°C (+350°C...+500°C)
TP473P wire Pt100 pointed probe. Tube Ø 4 mm, length 150 mm.. 4 wire cable 2m
long
Working range: -100°C...+400°C.
Accuracy: ±0.2°C (-100°C...+350°C) / ±0.4°C (+350°C...+400°C)
TP474C Thin film Pt100 contact probe. Tube Ø 4 mm, length 230 mm, contact surface
made of silver Ø 5 mm. 4 wire cable 2m long
Working range: -50°C...+400°C
Accuracy: ±0.2°C (-50°C...+350°C) / ±0.4°C (+350°C...+400°C)
Relative humidity and temperature combined probes
Typical characteristics of module of relative humidity and temperature probes
Relative Humidity
Sensor
Typical capacity @30%RH
Probe temperature working range -40°C...+150°C
Working range
Accuracy
Resolution
Temperature drift @20°C
%RH response time at
constant temperature
Temperature
Temperature sensor
Working range
Accuracy
Resolution
Temperature drift @20°C
Temperature sensor (HP572AC)
Working range
Accuracy
Resolution
Temperature drift @20°C
HP472AC RH% and temperature combined probe, dimensions Ø 26x170 mm.
Connecting cable: 2m long. Working range: -20°C...+80°C, 5...98% RH.
RH% accuracy ±2%
HP572AC RH% and K thermocouple temperature combined probe
Dimensions Ø 26x170 mm.
Working range: -20°C...+80°C, 5...98% RH.
UR% accuracy: ±2%
HP473AC RH% and temperature combined probe. Handle Ø 26x130 mm, Probe
Ø 14x110 mm. Connecting cable: 2m long.
Working range -20°C...80°C, 5...98% RH.
RH% accuracy: ±2%
HP474AC RH% and temperature combined probe. Handle Ø 26x130 mm, probe Ø
14x210 mm. Connecting cable: 2m long.
Working range: -40°C...+150°C, 5...98% RH.
RH% accuracy: ±2%
HP475AC RH% and temperature combined probe. Handle Ø 26x110 mm. Stainless
steel tube Ø12x560 mm. Terminal tip Ø 13,5x75 mm. Connecting cable:
2m long.
Working range: -40°C...+150°C, 5...98% RH.
RH% accuracy: ±2%
).
Peak time ≥ 5ms
Peak dead band ≤ 2% f.s.
Mk-33 capacitive
300pF±40pF
0 ... 100%R.H.
±1%UR in the range 20...90%RH
±2%UR in the range 10...99%RH
0.1%RH
0.02%RH/°C
10sec (10→80%RH; air speed=2m/s)
Pt100 (100Ω @ 0°C)
-50°C...+200°C.
±0.03°C
0.01°C
0.003%/°C
K thermocouple
-50°C...+200°C.
±0.5°C
0.05°C
0.02%/°C
°C accuracy: ±0.25°C.
Connecting cable: 2m long.
°C accuracy in: ±0.5°C.
°C accuracy: ±0.25°C.
°C accuracy:±0.25°C
°C accuracy: ±0.30°C
HP477DC %RH and temperature sword combined probe, handle Ø 26x110 mm.
Tube 18x4 mm, Length: 520 mm. Connecting cable: 2m.
Working range -40°C...+150°C, 5...98% RH.
.
RH% accuracy: ±2%
2
) that is
-2
Pressure Probes
PP472
Probe for measuring barometric pressure.
Working range: 600 ... 1100mbar
Accuracy @ 20°C: ±0.3mbar
TP704-705 Probes to couple to the SICRAM PP471 module to measure the abso-
lute, relative or differential pressure.
Maximum
Full Scale
Differential
overpres-
Pressure
Pressure
sure
NON- isolated
membrane
10.0 mbar 20.0 mbar TP705-10MBD
20.0 mbar 40.0 mbar TP705-20MBD
50.0 mbar 100 mbar TP705-50MBD
100 mbar 200 mbar TP705-100MBD
TP705-200MBD
200 mbar 400 mbar
TP705-500MBD
500 mbar 1000 mbar
TP705-1BD
1.00 bar
2.00 bar
TP705-2BD
2.00 bar
4.00 bar
5.00 bar
10.00 bar
10.0 bar
20.0 bar
20.0 bar
40.0 bar
50.0 bar
100.0 bar
100 bar
200 bar
200 bar
400 bar
500 bar
750 bar
PP473 S1...S8 Differential pressure probes
Working range
Maximum over-pressure S1,S2,S3=200mbar S4=300mbar
Accuracy @ 25°C
Temperature range
Fluid in contact with the membrane
Connection
5
°C accuracy: ±0.30°C
Resolution: 0.1mbar
Temperature range: -10 ... +60°C
Relative Pres-
sure (with
ACCURACY
ABSOLUTE
respect to the
from 20 to
Pressure
atmospheric
25°C
pressure)
isolated
isolated
membrane
membrane
0.50% FS
0.50% FS
0.50% FS
0.25% FS
0.12% FS
TP704-200MBGI
0.20% FS
0.12% FS
TP704-500MBGI
0.20% FS
0.12% FS
TP705-1BGI
0.20% FS
0.12% FS
TP704-2BGI
TP704-2BAI
0.40% FS
TP704-5BGI
TP704-5BAI
0.40% FS
TP704-10BGI
TP704-10BAI
0.40% FS
TP704-20BGI
TP704-20BAI
0.40% FS
TP704-50BGI
TP704-50BAI
0.40% FS
TP704-100BAI 0.40% FS
TP704-200BAI 0.40% FS
TP704-500BAI 0.40% FS
S1=f.s.10mbar,
S2=f.s.20mbar, S3=f.s.50mbar,
S4=f.s.100mbar,
S5=f.s.200mbar, S6=f.s.500mbar,
S7=f.s.1bar,
S8=f.s.2bar
S7=3bar
S8=6bar
±0.5%f.s. (10, 20,
±0.25%f.s.
50mbar)
(100mbar)
-10 ... +60°C
non-corrosive and dry gas or air
tube Ø 5mm
Working
Tempera-
Connection
ture
0...60°C Tube ∅5mm
0...60°C Tube ∅5mm
0...60°C Tube ∅5mm
0...60°C Tube ∅5mm
0...60°C Tube ∅5mm
0...80°C
1⁄4 BSP
0...60°C Tube ∅5mm
0...80°C
1⁄4 BSP
0...60°C Tube ∅5mm
0...80°C
1⁄4 BSP
0...60°C Tube ∅5mm
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
0...80°C
1⁄4 BSP
S5,S6=1bar
±0.12%f.s. (200,
500, 1000,