Glass
PTFE
5.3 Connecting the level switch
The GL 3 provides the option to install a model NS1/GL45 level
switch. The receptacle therefore has a GL45 connection.
DANGER
Explosion hazard
To use the GL 3 in an explosive atmosphere with level switch,
the level switch and controller must be ignition protection
class intrinsically safe "i". The use of non-intrinsically safe
components may result in ignition in explosive atmospheres.
Also note the specifications under EN 60079-14 in this respect.
The level switch must be screwed in tight, sealed with Teflon
tape or sealant/flat gasket!
Pay attention to leaks!
Please be sure to run the connection cable safely and without
tension. If possible, the cable between the level switch and
controller should NOT be run inside conduit with switch
cables, as these may cause interference.
Please refer to the separate operating and installation instruc-
tions included with the product for addition information on
connecting the level switch and controller.
6 Operation and Control
DANGER
The condensate may contain toxic or corrosive gases
The condensate may be hazardous to health.
a) Switch off the gas supply before performing mainten-
ance and, if necessary, flush the gas lines with air.
b) If necessary, ensure a safe gas discharge.
c) Protect yourself from toxic / acidic gasses when perform-
ing maintenance. Wear appropriate protective equip-
ment.
d) Only use adequately sized, durable transfer receptacles to
transfer the condensate.
The condensate trap level must be checked regularly and
emptied as necessary.
Proceed as follows:
• Remove the section of house from the equipment open-
ing and hold upward.
8
Bühler Technologies GmbH
• Remove the receptacle from the bracket.
• Open the drain valve and drain the condensate into the
transfer receptacle.
• Hold the transfer receptacle under the drain valve, or hold
the drain valve over the opening on the transfer recept-
acle.
• Once all the condensate has drained, close the drain valve.
• Return the receptacle to its bracket.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.
8 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.