MANUAL INSTRUCCIONES CODIGO 80199 REV B
Reading
Units
12.01
µS/cm
25.0ºC
CON
Status
Temperature
Sensor
Temp.
connector
sensor
Conductivity probe
Temperature sensor
J
.P. SELECTA s.a.
Abril/2010
4.1 Factory set parameters
Parameter:
Sensor constant:
ATC factor:
TDS factor:
ASD automatic stop:
5 Description
5.1 Display
•
Conductivity reading/TDS/Salinity
•
Units
•
Actual measurement temperature
•
Status indicator
5.2 Rear panel
•
Conductivity sensor connector.
•
Mini BNC connector for temperature sensor connection
•
Serial RS-232 connector
Output
RS-232
6 Operation
Once the equipment is unpacked, please visually check that all goods are OK. Keep
the packaging for some days.
6.1 Placing batteries
•
Place the batteries on the case, in the rear side of the pH-meter.
•
Pay attention polarity is correct.
•
If you are not going to use the unit for some time, please remove the batteries.
•
Replace the batteries when «LoBat» appears on the display.
6.2 Sensors connection
Connect the conductivity and the temperature sensor on the rear panel.
Do not move the unit with the sensors connected.
Note: The BNC connectors must be removed and fixed by turning 1/4 turn.
6.3 Panel
Switch on the pH-meter by pressing the «ON/OFF» key.
Wait until the equipment replies before pressing any other key.
Keys:
•
«ENTER/PRINT»
•
«MODE /MEAS»:
•
«
•
«
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
(Sujetas a modificaciones sin previo aviso)
Validate input.
Prints the reading.
Changes from MEASuring status to
MODE status and vice versa.
/C/T/S»:
Up. Change from conductivity, TDS or salinity.
/ SAVE»: Down. Stores the reading.
Pág..: 14
Valid range
Factory set
0.01; 0.1; 1; 5; 10
1.00
0.000 ... 0.040
0.020
0.40 ... 1.00
0.50
ON/OFF
ON
Tel 34 93 770 08 77 Fax 34 93 770 23 62