2. The display will show the first function, "F1
unit".
3. Pressing the TARE key will cycle through the
other functions.
4. Pressing ZERO key will allow you to set the
function.
5. It may be necessary to either use TARE key
set a value or using the M+ and PT key to
move the active digit and then using the
TARE key to increment a digit, followed by
the ZERO key to enter the value.
FUNCTION SETTING TABLE
FUNCTION
SUB-FUNCTION
F1 u nt
F2 dat
Scale will show y-n-d or d-n-y at first, this is date format (year-
month-date or date-month-year). Then you need set year,
month and date.
yy
mmdd
hhmm
F3 off
CLOCK
Bk
Off
F4 prt
0
1
2
3
4
5
13 OF 62
OPERATION CONTINUED
DESCRIPTION
Sets the displayed weight unit. Select kg/lb.
Set year: The display will show current year or 2000.
Enter new year.
Set date: The display will show current date or 01.01,
enter new date, format mm.dd (month/date).
Set time: The display will show current time or 00.00.
Enter new time, format hh.mm (hour/minute).
Set clock off or on.
Off: Don't use clock function.
On: Clock display after scale is on standby for
5 minutes.
Set the backlight.
On: Always on.
Auto: Press a key and weight is more than 20d,
backlight comes on automatically.
Off: Always off.
Set auto power off.
0: Don't use auto power off function.
3: Auto power off after standby 3 minutes.
5: Auto power off after standby 5 minutes.
15: Auto power off after standby 15 minutes.
30: Auto power off after standby 30 minutes.
Continues mode
Print formats
6. Use the "U" key to leave a parameter
unchanged.
7. For example when the display shows "F1
unit" press the ZERO key to begin.
The display will show "ON lb". Press the TARE
key to set the weighing unit to ON/OFF status.
8.
Press ZERO key to enter value, and the
scale will show the next weighing unit.
Use TARE key to set and ZERO key to
enter. When finished press U key to exit.
DEFAULT VALUE
kg/lb
2000/01/01
00:00
CLK off
Auto
of 0
0721 IH-4564