System Setup

PIP+
RFS
Scavenging Circuits (32009, 33008)
1.
Assemble the circuits per Figure 1 and install the mask on the circuit per Figure 2.
2.
Connect the reservoir bag and hoses per Figure 3 for RFS Circuits or Figure 5 for Accu-vac
Caution: The bag must be on the downspout, only one vacuum controller can be used at one time (RFS Circuits or Accu-vac Circuits).
3.
Adjust the flow on the RFS Circuit using knob
that it be checked periodically during the procedure as facility vacuum pressure can change from time to time.
4.
Place mask on patient per Figure 6.
5.
Adjust the flow on the Accu-vac Circuit using knob
red=blocked flow (check for kinked hose).
PIP+
Inline Circuits (32007, 33010)
1.
Assemble the circuits per Figure 7 and install the mask on the circuit per Figure 8.
2.
Connect the reservoir bag and corrugated hose per Figure 9.
Caution: The bag must be on the downspout, only one vacuum controller
3.
Connect the vacuum line to the vacuum source such as the chair
fitting T per Figure 11 after cutting off
4.
Place mask on patient per Figure 6.
5.
Adjust the flow on the controller
down after initial setting although it is recommended that it be checked periodically during the procedure as facility vacuum pressure
can change from time to time.
Gas Flow Operation
Digital Ultra
Flowmeters:
breathing rate of the patient by watching the reservoir bag. At the proper flow it will neither inflate to the max or deflate to the minimum.
Gradually increase %N
O until desired effect is reached. Use the Titration method when adding N
2
end of procedure, reduce %N
Ultra PC
Flowmeters:
Turn the unit on and adjust the Total Flow to match the breathing rate of the patient by watching the reservoir bag. 
At the proper flow it will neither inflate to the max or deflate to the minimum. Gradually increase %N
that the Oxygen level will need to be reduced by the amount of increase in N
when adding N
O to avoid negative patient reactions. At end of procedure, reduce %N
2
maintain proper flow. Run for 5 minutes to Flush N
Turn unit on. O
Ultra DC
Flowmeters:
ing the reservoir bag.  At the proper flow it will neither inflate to the max or deflate to the minimum. Gradually increase N
effect is reached. Oxygen flow will need to be decreased by the same amount of the N
the Titration method when adding N
flush N
O from patient's blood stream. The oxygen flow will have to be increased by the amount of N
2
Flow.
PIP+
Circuits can be reprocessed per 26613-INT. Do not exceed recommended temperatures.
UDI Product Tag included for product traceability purposes.
so that the needle is in the green zone of gauge
M
W
Q
.
N
per Figure 12 using knob
Turn on the unit. It should start at TOTAL FLOW of 5 lpm with 0 %N
O to 0 for five minutes to flush N
2
O from patient's blood stream.
2
flow should start at 3 lpm. Adjust Oxygen flow to match the breathing rate of the patient by watch-
2
O to avoid negative patient reactions. At the end of the procedure, reduce N
2
so that the light
K
is green, blue=low flow (check vacuum flow from pump),
N
can be used at one time.
per Figure 10 using installed fitting
V
R
so that the needle is in the green zone of gauge
O from patient's blood stream.
2
O to maintain the correct flow rate. Use the Titration method
2
Circuits.
L
per Figure 4. It is recommended
Q
or connect to wall outlet using
O. Adjust TOTAL FLOW to match the
2
O to avoid negative patient reactions. At
2
O until desired effect is reached. Note
2
O to 0 and increase O
flow by same amount to
2
2
O increase to maintain the proper Total Flow. Use
2
O decrease to maintain the same Total
2
EN
S
. Gauge can be laid
O until desired
2
O to 0 for five minutes to
2