compliance level above, Dual Pex should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting of relocating Dual Pex.
** Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the medical device.
Dual Pex is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of Dual Pex can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and Dual Pex as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
16. Warranty
Inspections
and
repairs
Your product is an accurate medical device, produced under maximum quality and test conditions.
To ensure an individualized follow-up and traceability of your product, please complete the warranty
registration
on-line
activate the after-sales procedures for your device so that Micro-Mega can give you full satisfaction.
Micro-Mega guarantees the battery for one year and the device for two years against any hidden manufacturing defect,
excluding faults due to improper use, inappropriate maintenance, accidental damage (fall, impact, etc.), attempted
intervention or repairs by personnel not approved by Micro-Mega, attempted product modification.
17. Statement
Non-contractual pictures. Micro-Mega reserves the right to modify the characteristics of its product without prior notice.
Imported in UE by Micro-Mega S.A
12 rue du Tunnel
25000 Besancon - France
Caretechion GmbH
Niederrheinstr. 71, 40474
Duesseldorf - Germany
8
Recommended separation distances between portable and mobile RF communications
equipment and the Dual Pex.
150 kHz to 80 MHz
d=1.2√
P
0.13
0.36
1.1
3.5
13
must
only
be
on
www.micro-mega.com
80 MHz to 800 MHz
d=1.2√
P
0.11
0.37
1.3
3.6
10
carried
out
by
an
website
within
10
Manufactured by Changzhou Sifary Medical
Technology Co., Ltd.
No. 99, Qingyang Road, Xuejia County, Xinbei
District, Changzhou City, 213000 Jiangsu - China
800 MHz to 2.5 GHz
d=2.3√
P
0.22
0.72
2.1
7.1
20
approved
dealer
or
days
of
your
purchase.
Year of CE marking: 2017
by
Micro-Mega.
This
will