Prodigy
If the control solution test result is out of range.
v
Probable Cause
Error in performing
the test.
Control Solution bottle
not shaken well.
Expired or contaminated
Control Solution.
Control Solution that is
too warm or too cold.
Test strip deterioration.
Meter malfunction.
Prodigy Diabetes Care, LLC
Talking Meter
What to Do
Read the instructions
thoroughly and repeat the
test again.
Shake the Control Solution
bottle vigorously and repeat
the test again.
Check the expiration date
or the discard date of the
Control Solution.
Control Solution, meter and
test strips should come to
room temperature (68 –77°F/
20–25°C) before testing.
Please repeat the test with
a new test strip.
Please contact Technical
Support at 1.800.243.2636
for assistance.
39
Owner's Manual
1.800.243.2636