SET UP INSTRUCTIONS (CONTINUED)
pressing Zero / Tare. Users will see dashes on the screen as the scale is finding
zero.
PU2 = E
Enabling this feature means the scale will zero itself when it is powered on. After
each use, the scale will clear the weight of the patient and set itself back to the
initial ZERO level it found on start-up. Health o meter® Professional recommends
keeping this setting as Enable. Choose the setting you prefer and press
CLEAR/REWEIGH to save your selection and exit the Options mode.
2. The scale will restart. The display will show "UEr", followed by a version number, and then
dashes. When the display shows "0.0" the scale is ready to use.
®

Everlock

®
The Everlock
feature allows the measuring unit to be permanently locked on the scale, disabling
the UNIT button. The Everlock
®
Everlock
.
Warning: This lock is permanent and cannot be reversed.
Notes:
 Press CLEAR/REWEIGH at any time to exit the Everlock
 Once Everlock
scrolling through the Options menu.
 There is no method to undo Everlock
1. While the scale is turned off, press and hold the ENTER button then press and release the
button. Hold the ENTER button until "AOF" appears on the display. Use the ▲ or ▼
buttons to advance the display screen until "ELC" appears. Press the ENTER button to
change the setting.
ELC =d
Default setting with Everlock
ELC = E
Engages Everlock
begin the locking procedure. If this is desired, follow the instructions below.
2. After Everlock
enter a password. Press the following keys in this sequence:
BMI, ENTER, ▼, ▲
3. If entered correctly, "ELC=L" (if scale was in LB mode) or "ELC=6" (if scale was in KG mode)
will appear on the display.
4. Use the ▼ and ▲ keys to toggle between (L) Pounds and (6) Kilograms. When the desired
unit is displayed, press ENTER and "PASS2" will appear on the display.
5. This will prompt the user to enter the second password. Press the following keys in this
sequence:
▲, ▼, ENTER, BMI
6. If entered correctly, the scale will display "ELC = L" or "ELC = 6," depending on the unit of
measurement that was selected. Everlock
P/N UM501KL_501KG_REV20201201_web
default setting is "disabled". Follow this procedure to engage
®
®
is engaged the Unt and UdP options will no longer be displayed when
®
. Once this option is selected, the display will prompt you to
is selected, "PASS1" will appear on the display. This prompts the user to
®
®
procedure.
®
without sending the scale back to the manufacturer.
®
disabled.
®
is now engaged.