Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Type BF appliance
It is compulsory to carefully read the instructions
before using this device
The device's casing is protected against
solids 12.5 mm or more in diameter, against
vertically falling drops of water and against
f i n g e r a c c e s s t o h a z a r d o u s p a r t s .
Switch On
Switch off (or battery on charge)
Direct current
Alternate current
Compliance levels according to EN 60601-1-2:2015 standard
- ESD immunity: 15 kV air, 8 kV contact (EN 61000-4-2)
- Burst immunity: 2 kV/100 kHz (EN 61000-4-4)
- Surge immunity (EN 61000-4-5): 1 kV common mode /2 kV differential mode
- Magnetic field (EN 61000-4-8): 30 A/m
- Immunity to rf currents in the 150 kHz-80 MHz range (EN 61000-4-6) 3 V modulation 80% 1 kHz
- RF emissions, CISPR 11: Class B
- Harmonics emissions, EN 61000-3-2: Class A
Field (V/m)
3
27
28
9
28
28
28
9

SYMBOLS USED

Electromagnetic Compatibility
Rf field immunity (EN 61000-4-3):
Frequency
80MHz 2700MHz
380MHz 390MHz
430MHz 470MHz
704MHz 787MHz
800MHz 960MHz
1700MHz 1990MHz
2400MHz 2570MHz
5100MHz 5800MHz
22
Do not use the unit when takig a bath or a
shower
Conforms to Directive 93/42/EEC for
medical devices
The device contains a hermetic lead
battery. It must be disposed of in
accordance with current regulations
on the disposal of toxic-harmful
waste
Sterilisation by ethyl oxide
Disposable
1kHz AM 80%
18Hz PM 50%
18Hz PM 50%
217Hz PM 50%
18Hz PM 50%
217Hz PM 50%
217Hz PM 50%
217Hz PM 50%
Modulation
Tabla de contenido
loading

Productos relacionados para LEM Aspimed 1.9

Tabla de contenido