N.B.: If using chemical disinfectants, follow the manufacturer's instructions exactly.
•
The cannula and the aspirated flow manual command are sterile, disposable products and must be replaced
after every application.
•
The antibacterial disposable filter must be replaced after every application.
•
Never leave the appliance in water or submerged; clean the external casing of the appliance using only a
damp cloth with detergent (non abrasive).
PROBLEMS
Excessive noise.
The unit switches on but does not aspirate. - Damaged pump.
The vacuum rate cannot be regulated. Damage to the internal hydraulic system
When the appliance is switched on, the
protection fuse always trips.
The vacuum gauge does not work. Liquid penetrating the pneumatic circuit. Send to the assistance circuit.
Note: if you experience faults or malfunctioning problems different to those listed above, always and
exclusively contact authorised assistance centres.
Professional portable medical device for secretion removal, suitable for fixed and transportable use.
Complete with flame retardant ABS V0 casing and piston-cylinder electrical compressor lubricated for life.
Risk class under Directive 93/42/EEC:
Suction class:
Voltage:
Absorption:
Adjustable vacuum level:
Vacuum gauge precision class:
Air flow:
Temporary use:
Dimensions:
Weight:
Noise level:
Multi-voltage switching battery charger:
Internal battery:
Battery charge duration:
Fuse for cigarette lighter power supply cord:
Operating conditions:
Temperature:
Air humidity:
Storage conditions:
Temperature:
Air humidity:
Operating-storage conditions:
CLEANING AND DISINFECTION OPERATIONS
PROBLEMS, CAUSES AND SOLUTIONS
Damaged pump or blockages in the
internal aspiration circuit.
- Vacuum regulator fully open. Connection
tubes disconnected and/or badly
connected, broken connection tubes.
Container not in a vertical position, full, or
defective overflow valve. Possible
blockage of the hydraulic circuit inside the
unit.
or blockage of the connection tubes to
the aspiration unit.
Pump probably damaged or in shortcircuit. Send to the assistance circuit.
TECHNICAL CHARACTERISTICS
POSSIBLE CAUSES
IIa
High vacuum / High flow
12VDC
3A
from 0 to –0.85 bar (-85KPa)
2.5% under UNI EN 837
30 lt/min operating without restrictions +/– 10 %
max. 45 minutes
38 (length) x 13.5 (width) x 26 (height) cm (1000ml Vessel)
38 (length) x 13.5 (width) x 35 (height) cm (2000ml Vessel)
4.5 kg approx (1000ml Vessel)
4.7 kg approx (2000ml Vessel)
55dBA
PRI: 100 ÷ 240V~ 50 – 60Hz
SEC: 14VDC 4.28A
12VDC 4Ah Lead, hermetic
45 minutes at maximum suction power
F6.3A - 250V delayed fuse - dimensions Ø 6.3 x 30 mm
min. 0° C; max 40° C
min. 10 %; max 95 %
min. –10° C; max 50° C
min. 10 %; max 95 %
min. 690 hPa; max 1060 hPa
21
SOLUTIONS
Send to the assistance circuit.
- Send to the assistance circuit.
- Check the position of the vacuum
regulator. Check the connections and the
integrity of the tubes. Position the
container in a vertical position, check the
overflow valve (blocked) and/or replace
the silicon tubes.
Send to the assistance circuit.
%