The scale is calibrated before leaving the manufacturer's
facility. Vibrations in shipping may cause the scale to
fall out of calibration. Uline recommends re-verifying
calibration in the field before first use. You must use a
precision calibration weight to properly calibrate the
scale. The calibration weight must be at least 1/3 of
the scale's capacity and cannot exceed the scale's
capacity.
NOTE: Calibration can be done in English
system weight (lb.) or Metric system weight (kg).
CALIBRATION WEIGHT (lb.)
Model #
1/3 WEIGHT (Minimum)
H-1654
2 lb.
H-1653
2 lb.
H-1649
4 lb.
H-1650
10 lb.
H-1651
20 lb.
CALIBRATION WEIGHT (kg.)
Model #
1/3 WEIGHT (Minimum)
H-1654
1 kg.
H-1653
1 kg.
H-1649
2 kg.
H-1650
5 kg.
H-1651
10 kg.
NOTE: Calibration weights are sold separately.
See uline.com for details.
SINGLE POINT CALIBRATION (SPAN)
Single point calibration uses two calibration points: zero
and any weight between 1/3 of the scale's capacity
and full weight of the scale capacity. Refer to the
calibration weights charts above.
1. Turn on the scale. In weighing mode, press the
UNIT key
Unit
until the desired weighing unit
appears – lb or kg.
2. Turn off the scale. Turn on the scale. During the
self-test, press the % key
simultaneously and release. UnloAd will be shown on
the display.
UnLoAd
3. Remove all weight from the platform.
4. Once the stable indicator O is shown on the display,
press the ZERO key
used will be shown on the display.
PAGE 7 OF 27
FULL CAPACITY (Maximum)
6 lb.
6 lb.
12 lb.
30 lb.
60 lb.
FULL CAPACITY (Maximum)
3 kg.
3 kg.
6 kg.
15 kg.
30 kg.
%
and TARE key
. The last calibration weight
Zero
CALIBRATION
5. Use the %
enter the calibration weight you are using.
6. Press the ZERO key
display.
LoAd
7.
Place the calibration weight on the platform.
8. Once the stable indicator O is shown on the display,
press the ZERO key
9. Scale will enter a self-test mode. Remove weight
before self-test mode is complete. After completing
the self-test, the scale will enter the weighing mode.
MULTI-POINT CALIBRATION (LINEAR)
Multi-point calibration uses three calibration points: zero,
1/3 and full weight capacity of the scale. Refer to the
calibration weights charts above.
1. Turn on the scale. In weighing mode, press the
UNIT key
– lb or kg.
2. Turn off the scale. Turn on the scale. During the
self-test, press the % key
simultaneously and release. UnloAd will be shown on
the display.
UnLoAd
3. Remove all weight from the platform.
4. Once the stable indicator O is shown on the display,
press the ZERO key
will be shown on the display.
5. Place the 1/3 calibration weight on the platform.
6. Once the stable indicator O is shown on the
Tare
display, press the ZERO key
calibration weight will be shown on the display.
7.
Place the full capacity calibration weight on the
platform.
8. Once the stable indicator O is shown on the display,
press the ZERO key
9. Scale will enter a self-test mode. Remove weight
before self-test mode is complete. After completing
the self-test, the scale will enter the weighing mode.
%
Limit
, LIMIT
and TARE
. LoAd will be shown on the
Zero
.
Zero
until the desired weighing unit appears
Unit
and LIMIT key
%
. The 1/3 calibration weight
Zero
. The full capacity
Zero
.
Zero
Tare
keys to
Limit
0621 IH-1649

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

H-1650H-1651H-1653H-1654

Tabla de contenido