4–2 Pairing
To use the Link Module, you must first pair the Link Module and the Surgic Pro2.
Surgic Pro2 operation
3,5 2
Turn on the main power switch of Surgic
1
Pro2.
Press and hold the Setting key for 2
2
seconds or more to display "BLE SET".
Press the Enter key.
3
Press the TORQUE key and select "LK".
4
Press the Enter key.
5
Link Module operation
Turn on the VarioSurg 3. The LED of the
6
Link Module blinks white.
Press and hold the button on the Link
7
Module.
4
When the LED turns blue, release the
8
button.
Pairing is successful when the Surgic
9
Pro2 beeps and displays "OK". After
pairing, the Surgic Pro2 is automatically
connected to the Link Module. During
connection, the LED of the Link Module
blinks green and the Bluetooth mark
blinks on the left of the "VARIOSURG"
icon on the Surgic Pro2 display.
10
Press the Cancel key of the Surgic Pro2
twice to end setting mode.
5