91
Guidance and manufacturer's declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS that are
not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz
to 80 MHz
Radiated RF
3 V/m
IEC61000-4-3
80 MHz
to 2.5 GHz
Compliance
Electromagnetic environment - guidance
level
3 Vrms
Portable and mobile RF communications equipment should be used
no closer to any part of the device, including cables, than the recom-
mended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
3 V/m
3.5
d=
V
1
3.5
d=
E
1
7
d=
E
1
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the rec-
ommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an elec-
tromagnetic site survey,
each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
ENGLISH
√ P
√ P
80 MHz to 800 MHz
√ P
800 MHz to 2.5 GHz
a
should be less than the compliance level in
b
Tabla de contenido