Do not use the device after extreme temperature fluctuations!
S
O
WITCH
N
After switching on the device, a welcome screen will be displayed while the device performs
a self-test. At the end of the self-test the unit enters the "home menu" and is ready for use.
H
M
OME
ENU
From the Home menu, user can start a
treatment by selecting
1) A preset clinical protocol, or by
2) Retrieving a previously stored
favorite or
3) Setting a protocol manually.
User can also set the desired system
settings from home menu.
M
O
ODE OF
PERATION
The device incorporates smart mode or conventional mode of operation.
In conventional mode (=smart mode off), peak power, pulse repetition rate, pulse duration
and treatment time can be set within the required range for the electrodes that can be
connected to the device. All these parameters are shown in the treatments screen. The mean
power that is the result from these parameters will be shown too.
In smart mode, all one needs to set is the mean power. Based on the best available evidence,
Enraf-Nonius has developed an algorithm that will select the appropriate Pulse repetition
frequency, Pulse duration and Peak power to go with your selected Mean power. The unique
incorporated slider bar will even help you decide on picking the right Mean power based on
the character of the indication (Acute to Chronic scale) or the need for a non-thermal or
thermal treatment.
To select the mode of operation, go to system settings. By default, it is set to "On".
EN109-1670750-46 IFU