Setting Memory Locations; Setting Mixing Time; Installing Capsules; Starting The Mixing Cycle - Dentsply Sirona ProMix 400 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

SETTING MEMORY LOCATIONS:

The ProMix® Amalgamator memory locations (M1 & M2) can be programmed to suit the
two most commonly used amalgamation cycle times and speed settings of the individ-
ual user (corresponding to the manufacturer's specification of the material/amalgam
being mixed).
a) Select the desired speed by pressing the appropriate key pad (verify the asso-
ciated LED illuminates).
b) Select the desired run time by depressing and holding the up or down arrow
button/pad (to scroll through time) until the desired time is displayed on the
LCD.
c) Store the setting by depressing and holding the desired memory location (M1
or M2) button/pad for longer than 3 seconds (the LCD will display "00" for 0.5
second before the desired memory setting will reappear indicating the end of
the storage cycle).
SETTING THE MIXING TIME:
NOTE: Refer to the encapsulated material manufacturer's specification to
determine the proper mixing time and speed.
If the material/amalgam required mixing time and speed is the same as one of the two
(2) stored memory locations (M1 or M2) simply depress that desired Memory key pad
and proceed to "Installing Capsules".
If not, proceed as follows:
a) Select the recommended speed by pressing the appropriate speed
button/pad:
i) High speed (rabbit)
ii) Low speed (turtle)
b) Select the recommended mixing time (to a maximum 30 seconds) by:
i) pressing the appropriate Up or Down arrow button/pad until the
recommended mixing time is displayed on the LCD; or holding down the
Up or Down arrow button/pad automatically scrolls through the mixing
time in one (1) second intervals.
8

INSTALLING CAPSULES:

The ProMix® Amalgamator capsule holder is designed for easy use and to
automatically retain the capsule when installed properly (refer to Figure 2, inside front
cover).
a) Lift the ProMix® Amalgamator cover to gain access to capsule holder.
b) Open the capsule holder by pressing the Arm Button with your thumb while
holding the Arm Body Tab with your fingers to keep the assembly stable.
c) Activate the capsule (if necessary, per material manufacturer's instructions).
d) Place the capsule into the opened jaws of the Arm assembly.
NOTE: Capsules with nozzles should be placed to the left (such that the noz-
zle fits in the groove of the capsule holder).
To assure consistent mixes, capsules of the same brand should be oriented in the same
position (e.g. top to left) each time.
e) Release the Arm Button, which automatically retains the capsule.
f) Close the ProMix® Amalgamator cover.
STARTING THE MIXING CYCLE/UNIT OPERATION:
a) To start the mixing cycle, simply press the Start/Stop button.
b) The LCD display will count down to "00" from the selected run time in one (1)
second increments.
c) The ProMix® Amalgamator will automatically Stop at the end of the run cycle.
d) The LCD display will return to the previously selected run time.
STOPPING/INTERRUPTING THE CYCLE:
NOTE: Interrupting/halting the mixing cycle in any of the following
manners may make the material being processed unusable (refer to
the material/ amalgam manufacturers specification).
9