5.0 SYSTEM OPERATING INSTRUCTIONS
5.1
Pre-procedure preparation
1. Choose the appropriate set-up per Section 4.2 for the intended procedure.
2. Verify the power cord is plugged into the unit and an appropriate electrical receptacle.
3. Verify the footswitch cord is attached to the unit. Uncoil the footswitch cord and place the
footswitch near the operator's station.
4. Turn on the power switch. Note: On is the "I" position. Off is the "O" position.
5. Rotate motor speed control knob slightly to verify that the motor speed varies.
6. Press the footswitch to ensure that the unit starts and stops using the footswitch.
7. Turn off the power switch until the smoke evacuator is required.
5.2
Operative Use
Warning: The smoke evacuator produces a strong vacuum. Properly adjust the suction control
and the position of the inlet end of the suction tubing or wand to prevent injury to the
patient or inadvertent damage to surgical materials.
Warning: Some cannulas, valves, or other instruments combined with suction may cause rapid
reduction of the pneumoperitoneum in laparoscopic procedures.
1. Move the SurgiFresh
2. Attach tubing per the instructions in Section 4.2.
3. Turn on the power switch.
4. The footswitch may be pressed by personnel in the operative area to start and stop the
suction as needed during the procedure. Place the footswitch in a convenient location.
5. Rotate the motor speed control knob clockwise to increase air flow (suction) and
counterclockwise to decrease air flow. Stop at desired setting.
6. Verify adequate air flow is set by inquiring with the surgeon. Adjust as necessary with the
motor speed control knob. Turning the control knob completely clockwise places the unit in
turbo mode.
7. When the procedure is finished, turn off the power switch.
8. Collect the footswitch and electrical cord.
9. Follow cleanup instructions per Section 5.4.
5.3
Maximizing Use Suggestions
1. Air flow
a.
The distance of the tubing from the plume source affects capture and removal of the
plume. The tubing end should be within several inches of the procedure site.
b.
Larger tubing will permit a higher volume of flow for better plume removal. Do not occlude
tubing.
2.
Noise
a.
Lower speed equals less noise.
b.
Larger tubing equals less noise. Do not occlude tubing.
c.
Use the footswitch to turn the suction on and off as it is needed during the procedure.
3.
Incidental Fluids
a.
Do not allow a suction canister to overfill.
b.
Replace filters immediately if contaminated by fluids.
c.
The SurgiFresh
OPERATING INSTRUCTIONS
®
Turbo into position near the procedure site.
®
Turbo should not be used as a primary fluid suction system.
7