4.0 SYSTEM SET-UP
4.1 Installation
®
The SurgiFresh
Turbo with manual will arrive in a single carton. Inspect the carton and the
product for any scratches, dents or damage which may have occurred during shipment. If any damage is noted,
follow the instructions per the Warranty section. Smoke evacuation disposables such as filters, tubing and
accessories may be shipped separately. The same shipping damage instructions apply to
these items.
After removal from the carton, the SurgiFresh
Note: Save carton and packing for future use.
1.
Connect the footswitch tube to the side tubing port on the smoke evacuator. Place the footswitch control
in the footswitch mode. If the footswitch is not desired for use, override the footswitch by placing the
footswitch control in the normal mode.
2.
Install an ULPA filter in the center core of the unit. Note the air flow direction arrow on the ULPA filter.
The arrow should be pointing downward. Be sure there are no obstructions and the filter
seats completely by pushing all the way down.
Note:
Fill out the date sticker (included with filter) and apply it to the top of the unit. The normal replacement is
90 days from the date of installation. Early replacement is possible due to heavy usage as indicated by the ULPA
filter test or odor.
3.
Connect the power cord to the unit and an appropriate electrical receptacle.
4.
Perform the ULPA filter test per the procedure in Section 6.2.
Note:
The ULPA filter test is performed with no pre-filter or tubing in place. After testing is satisfactorily
completed, turn the motor speed knob to approximately ¼ speed to ensure that the unit will not be
activated at high speed.
5.
Install a pre-filter to the top of the ULPA filter.
Caution: Be sure that both the ULPA filter and the pre-filter are securely seated to ensure proper
suction flow.
6.
Determine the type of procedure, select the appropriate tubing and accessories, and prepare the unit for
use as described in Section 4.2
Warning: Sterile tubing should be handled using your institution's sterile protocol.
4.2 Typical Setup Instructions
1.
OPEN PROCEDURES -- EVACUATING DRY PLUME / SMOKE
a. Ensure an ULPA filter and applicable pre-filter are installed per section 4.1
b. Connect an adapter, if required, and/or tubing appropriate for the procedure to the pre-filter
OPERATING INSTRUCTIONS
®
Turbo is set up as follows: