FIELD CALIBRATION TEST

This laser level left the factory fully calibrated. Ironside recommends
the user check the accuracy of the laser periodically, especially if the
unit falls or is mishandled.
1. Check the height accuracy of the cross created by the lateral (#5)
and the horizontal lines.
2. Check the height accuracy of the cross created by the
longitudinal (#4) and the horizontal lines.
3. Check the accuracy of the forward vertical beam.
4. Check the accuracy of the side vertical beam.
5. Check the perpendicularity between the 2 vertical beams.
1.Checking the Height Accuracy of the cross created by the
lateral (#5) Horizontal lines.
(Up and down deviation)
1) Set up the laser on a table or on the floor between 3 walls A, B
and C. The distance between A and B should be approximately
5 meters.
2) Position the laser level approximately 0.5 meters from wall A and
2 meters from wall C.
3) Unlock the pendulum and press the Beam Selector (a), twice to
project all 3 lasers beams.
4) Direct the cross lines of the horizontal beam and the side vertical
beam (# 5) towards wall A.
5) Mark on wall A the center of the cross beams as a1, and on wall C
mark the center of the cross beam as c1 (see figure # 1).