General Features
Power Supply 100÷240 Vac
•
Display 4 digits with 7 segments
•
Bnc connection for probe Redox
•
Switch 3 position Priming, ON, Stand by
•
Peristaltic pump flowrate and pressure look at label on product
•
Measure Range Redox: -990÷990 mV (resolution: 10 mV)
•
Dosage method ON-OFF ( Set Point adjustable)
•
Calibration method Redox 1 Point (465 mV)
•
Switch position
:
1) Priming position "the pump running" (Switch position MOM)
2) Normal position "the instrument measure Redox and dosing. (Switch position ON)
Prog
Keep Press
key for 5 seconds to run calibration menu:
Redox probe: the pump displays C465 press
Displays quality probe ex: C100
Keep Press
when you are ready to leave a
Value to set from -990 until +990 mV increase step by step 10 mV.
3) Stand by position "The pump is stop" (Switch position OFF)
Prog
Keep Press
key for 3 seconds to run Setup menu:
Redox setup menu:
Adjust dosage method High ("Hi") or Low ("Lo").
NIKE Redox
key until flashing Setpoint value press again to increase the value
key after 3 seconds the unit save the value and flashing mem.
code: 0000137812 rev. 1.0
NIKE Redox
Prog
key and wait 60 seconds