1 Warning and safety information

1
Warning and safety information
Warning and safety information
Symbols used
Proper use
General safety information
Modifications to the system
Disturbance of electronic equipment worn by the
patient.
Radiation protection
4
To prevent personal injury and damage to property, ple-
ase observe the warning and safety information provi-
ded in these operating instructions. They are highlighted
with a warning symbol and with the word "CAUTION".
Observe accompanying documents (on name
plate)
This unit has been designed for use in creating intraoral
radiographic exposures.
The unit must not be used in areas where there is a risk
of explosion.
Improper use or stress (e.g. hard contact of the tube
assembly / support arm with another object, overexten-
sion of the support arm) may cause damage to the X-ray
equipment or the support arm.
In such cases, an inspection of the X-ray system must be
performed by a service engineer.
Patient X-ray exposures may be performed only when
the unit is functioning flawlessly, this means, e.g. even,
jolt-free and silent movement of the support arm.
The patient must not be left unattended at the unit.
Modifications to this system which could impair the
safety of operators, patients or third persons are prohib-
ited by law!
For reasons of product safety, this product may be
operated only with original Sirona accessories or
third-party accessories expressly approved by Sirona.
The user is responsible for dangers resulting from the
use of non-approved accessories.
To prevent the malfunction of electronic devices and
loss of stored data in connection with radio controlled
clocks, telephone cards, etc. all such items must be
removed from the patient prior to the X-ray exposure.
The applicable radiation protection regulations must be
observed. The prescribed radiation protection accesso-
ries must be used. If the release button is connected with
a cable, the operator must leave the vicintiy of the X-ray
tube unit before releasing an exposure.
No one except the patient may remain in the exposure
room. In exceptional cases a third party who is not a
member of the practice staff may assist the patient.
Visual contact with the patient and the unit must be
maintained during every X-ray exposure.
In case of malfunction, interrupt the exposure immedia-
tely by letting go of the release button.