I1O?
RANGE1 <nrf>
RANGE1?
DELTAV1 <nrf>
DELTAI1 <nrf>
DELTAV1?
DELTAI1?
INCV1
INCV1V
DECV1
DECV1V
INCI1
DECI1
OP1 <nrf>
SENSE1 <nrf>
TRIPRST
LOCAL
LSR1?
LSE1 <nrf>
LSE1?
SAV1 <nrf>
RCL1 <nrf>
System and Status Commands
*RST
EER?
QER?
*CLS
*ESE <nrf>
*ESE?
32
return the output readback current – response is <nr2>A<rmt> where <nr2> is in
Amps.
set the voltage range to <nrf> where <nrf> has the following meaning:
QL355P:
0=15V(5A), 1=35V(3A), 2=35V(500mA)
QL564P:
0=25V(4A), 1=56V(2A), 2=56V(500mA)
return the set voltage range – response is R1 <nr1><rmt> where <nr1> has the
following meaning:
QL355P:
0=15V(5A), 1=35V(3A), 2=35V(500mA)
QL564P:
0=25V(4A), 1=56V(2A), 2=56V(500mA)
set the output voltage step size to <nrf> Volts
set the output current step size to <nrf> Amps
return the output voltage step size – response is DELTAV1 <nr2><rmt>
where <nr2> is in Volts.
return the output current step size – response is DELTAI1 <nr2><rmt>
where <nr2> is in Amps.
increment the output voltage by step size
increment the output voltage by step size with verify
decrement the output voltage by step size
decrement output voltage by step size with verify
increment the output current limit by step size
decrement the output current limit by step size
set output on/off where <nrf> has the following meaning: 0=OFF, 1=ON
set output sense mode where <nrf> has the following meaning: 0=local, 1=remote
attempt to clear all trip conditions
go to local
query and clear LSR1, the limit status register − response is <nr1><rmt>
set the value of LSE1, the limit status enable register, to <nrf>
return the value of LSE1, the limit status enable register - response is <nr1><rmt>
save the current PSU set up to the set-up store specified by <nrf>
where <nrf> can be 0-9
recall a set up for the PSU from the set-up store specified by <nrf>
where <nrf> can be 0-9
Resets the instrument to the factory default settings − (see Factory Defaults section)
with the exception of all remote interface settings.
Query and clear Execution Error Register. The response format is nr1<rmt>.
Query and clear Query Error Register. The response format is nr1<rmt>
Clear Status. Clears the Standard Event Status Register, Query Error Register and
Execution Error Register. This indirectly clears the Status Byte Register.
Set the Standard Event Status Enable Register to the value of <nrf>.
Returns the value in the Standard Event Status Enable Register in <nr1> numeric
format. The syntax of the response is <nr1><rmt>