Please plug the product in socket securely and reliably to avoid leakage.
A separate dedicated socket must be used and grounded reliably. Do not lengthen
the power cord. If it is really necessary to lengthen it, be sure to use a copper
conductor of 2.5 square meters or more. The cross‐section of the copper conductor
in the wall connected to the power socket must be 4 square meters or more, in order
to avoid fever or fire.
Do not use power supplies other than those specified in the design of the equipment
to avoid overheating, short circuits, and other faults. Connecting a 110 V rated
voltage product to a 220 V power supply may cause overheating or burn out the
equipment. The machine uses a 220V/50HZ AC power supply. If the voltage is too low
or too high, an appropriate automatic voltage regulator must be used.
Please keep the product stable and avoid shaking.
Do not place the product in an area exposed to sun or rain, to prevent dangerous
accidents such as short circuit or overheating.
Do not tilt or lie on the side of the device. Do not hit the device body. The device is
equipped with a refrigeration system. The roll or impact is easily damaged.
Please place the product in a dry, dust‐free environment to avoid overheating, short
circuit, etc.
If unexpected sound, odour, smoke, etc. occurs when the power is turned on, unplug
the product and contact the manufacturer or supplier.
Please place the product in a dry and ventilated environment to ensure that the
device vents and instrument surfaces are not obstructed or blocked by walls or other
objects; do not use in a poorly ventilated environment to prevent damage due to
heat release from the device.
It is forbidden to dismantle and transform this equipment without authorization so
as to avoid potential safety hazards. In this case, manufacturer will not assume any
responsibility for quality accidents.
It is forbidden to put flammable and explosive dangerous goods and strong corrosive
acids, alkalis and other items that are not suitable for medical refrigerators.
When storing toxic, hazardous or radioactive materials, use a medical refrigerator in
a safe area. If used improperly, it may cause harm to human health or the
environment.
Do not insert metal objects such as nails or iron wires into any openings or gaps in
the medical cooler or any outlets, otherwise electric shock or injury may occur due to
accidental contact of the objects with moving parts.
In order to ensure the normal operation of the machine and ventilation and heat
dissipation, the back of the box and the left and right sides are at least 30cm away
from the wall. Please do not block the inlet and outlet with obstacles!