Manuale del d'uso ed installazione – WATER OK
A- Check the pressure in the water main using a pressure gauge. It must be at least 2 bar but no higher than 5 bar.
B- Fit a cock on the water main, with a connection of the type used for washing machines.
C- Fix the appliance in a position where the cartridges can be easily replaced.
D- Connect union 6 to the washing machine-type cock using the supplied pipe.
E- Connect union 4 to the autoclave using the hose pipe.
F- Connect tap 3 to the autoclave and close the safety fastener.
G- Set the autoclave program to the automatic mode (see instruction manual).
H- Connect the supplied transformer 2 to a 230V power socket.
I-Connect transformer 2 straight to Water Ok.
L- Using the supplied wrench, make sure that the cartridge covers are well tightened.
M- Make sure that everything operates correctly.
FAULTS
Difficulty in filling, or non-filling
alarm
The autoclave fails to fill
Green and red indicator light on
Red indicator on
No indicator light on
Del 03-01-2007 - D:\REVERBERI MEDICAL DIVISION\MANUALI\WATER OK.doc
WATER OK
INSTALLATION INSTRUCTIONS
CAUSES
Clogged filters.
Check the pressure in the water main.
Burnt-out solenoid valve coil.
The appliance fails to receive the command from
the autoclave.
Cartridges becoming exhausted.
Cartridges exhausted.
Transformer faulty or badly connected.
(GB)
Replace the filters.
Raise the pressure of clean the supplied
pipe filter.
Replace the coil or solenoid valve.
Check the autoclave socket (check the
fuse).
Check when only the red indicator light
remains on.
Replace the cartridges.
Check the connection or replace the
transformer.
REMEDIES
Pag. 4 di 7