3.0 Getting Started
3.1 Unpacking
Upon unpacking the Dry Bath, inspect for damages. Shipping damage is the responsibility of the carrier.
Open the
AccuBlock Dry Bath
•
Dry Bath unit
•
Data logger software
•
Line cord
If there are any items missing, damaged, or not according to your order, please contact your distributor or
sales representative immediately.
3.2 Initial Operation
Select a location that is dry and not subject to drafts or moving air from heating or air conditioning
vents, or air blown by other equipment. Place the unit on a flat, preferably non-flammable surface. Allow
sufficient room around the unit for access and cooling. Six inches minimum on all sides is suggested.
Plug the unit into a properly grounded outlet. Using the lifter, insert the block(s) into the well. The unit is
now ready for use.
4.0 Controls
4.1 Keypad Function
•
On/Off rocker switch. Located on back of unit. Turns primary power on and off
•
"Start/Stop" LED, red. Used to activate or stop the unit. Illuminates when unit is in heating mode, off in
temperature set mode
•
"Heating" LED, red. Illuminates when unit is in heating mode and unit is actually apply heat to the block.
This LED is on continuously during heat-up and cycles and off when the unit is at set temperature.
•
"Time" LED, green - Used to set or select time mode. This LED light indicates running Time Mode.
•
"TEMP-UP arrow key. Raises set temperature when unit is in set mode.
•
"TEMP-DOWN" arrow key. Lowers set temperature when unit is in set mode
•
"Timer-UP" arrow key. Raises set time when unit is in set mode.
•
"Timer-DOWN" arrow key. Lowers set-time when unit is in set mode.
•
To enter continuous mode, set the time to 00:00, hold the "Timer-DOWN" arrow key while pressing the
"Start/Stop" key. Unit will display "Hold".
2 | AccuBlock Digital Dry Bath
package and confirm that all items are included:
•
Block Lifter
•
Lid
•
USB cable