LITTLE GUY QUICK DISCONNECT
Thank you for selecting the Athena Champion Little Guy or Little Pro Handpiece.
Athena Champion products are engineered and manufactured to provide exceptional
performance, reliability and precision. Please take a few minutes to thoroughly read the
instructions for use to obtain peak performance and longevity from your new handpiece.
Important: The Food and Drug Administration and Centers for Disease Control and
Prevention strongly recommend that handpieces be resterilized after each patient use.
• Do not operate above 45 PSI.
• Operate only when the forward/reverse ring is in the full forward or full reverse
position (Little Guy only).
• Do not change rotational direction (turn forward or reverse ring) when the handpiece
is operating (Little Guy only).
• Lubricate and maintain your motor in accordance with the instructions. Proper care
and maintenance will extend the life of your Little Guy or Little Pro handpiece.
H
Prior to connecting the handpiece, check that all tubing supplying the handpiece hose
is properly connected with the dental unit. Make sure there are no restrictions in any of
the lines. Before connecting the motor, place one or two drops of Champ-Lube
the drive air port of the handpiece. Briefly operate the handpiece until excess lubricant
is expelled. Attach the threaded connector on the tubing set to the mating connector on
the handpiece. Ensure the small metal tubes entering from the rear of the handpiece are
properly nested in the matching holes of the tubing fitting and the gasket is in place. An
air leak will be apparent if the motor is improperly connected and the handpiece will not
run at the correct RPM.