Important Control Solution Information
FreeStyle® Control Solution is good for three months after opening
the bottle or until the expiration date printed on the label, whichever
comes first.
Important: Count forward three months from the date you open a new
bottle of control solution. This is your discard date. Write this date on
the side of the control solution bottle.
Example: You open the control solution on January 15th. Count
forward three months to April 15th. This is the discard date that you
record on the bottle.
•
Use only FreeStyle Control Solution with your FreeStyle Lite System.
•
Replace the cap on the control solution bottle immediately after
using.
•
Do not use control solution past the expiration date.
•
Do not add water or any liquid to control solution.
•
Control solution tests are specified to be accurate only when used
between 59° to 104° F (15° to 40° C).
•
Control solution test results should fall within the range printed on
the test strip vial label.
Important: The control solution range is a target range for control
solution only. It is not a target range for your blood glucose level.
•
If control solution results are out of this range, repeat the test.
•
If your control solution test results continue to fall outside the range
printed on the test strip vial, do not use the FreeStyle Lite System to
test your blood glucose. Contact Customer Care.
Caution: Results from FreeStyle Control Solution tests do not
reflect your blood glucose level.