English
8.0
Technical Information
General Specifications:
Product Type:
281
No. of Channels:
2
Waveform:
Symmetric Bi-Phasic
Environmental Specifications:
Operation:
Temperature 0° to 35° C
Humidity
Storage:
Temperature 0° to 55° C
Humidity
Physical Specifications:
Unit Dimensions:
Weight
• Unit
• Unit with battery
Nominal output voltage / power
Parameter
Output RMS voltage (RMSV)
Output RMS current (RMSA)
Output frequency
DC Component
Pulse Width
Current Intensity Range (per pulse)
20 to 65 % RH
10 to 90 % RH
105 x 71 x 31mm
93 g
140 g
XP units are products of BMR Ltd, Parkmore Business Park
West, Galway, Ireland.
Safety Features
Safe start: The intensity is set automatically to zero when
the unit is turned on.
Multiplexing: Pulse delivery to each channel is off-set so
that only one channel is energised at any instant. This
ensures there is no interaction between the electrodes of
each channel.
Electrode area less than 6.5 cm
in excess of 2m/cm
at maximum intensity. If in doubt, contact
2
neurotech or your clinician.
500
Ω
1k
Ω
7.5 V
12.3 V
15 mA
12.3 mA
4-99 Hz
4-99 Hz
0 C
0 C
80 – 400 µs
80 – 400 µs
0 – 75 mA
0 – 75 mA
can cause current densities
2
1.5k
Ω
13.7 V
9.2 mA
4-99 Hz
0 C
80 – 400 µs
0 – 75 mA
15