Connect the short silicon tube 1, with antibacterial filter 2, to the suction connector 3. The other tube, with one end connected to the filter must
be connected with the other end to the jar's connector 4 where has been fixed the red float (security float).
•
When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on the jar)
to avoid liquid penetration inside the device. The device must be used on a plan of horizontal operation.
•
Connect the long silicon tube 5 to the other jar's lid connector 6
•
Connect the other end of the long silicon tube 5 to the probe plastic connector 7
•
Connect the power cord to the device than connect the plug to the electrical mains supply.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the functionality
vacuum) then rescrew the lid on the jar correctly.
•
Push switch 8 on position I to start suction
•
Once finished push switch on O position and unplug
•
To extract the accessories and start with cleaning it.
WARNING: Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the "VACUUM".
assembling
Filter
Suction Pump Inlet
The dual use version is available on request. Continuous use can be enabled by activating the illuminated switch.
Alternatively, the pedal control can be used for aspirating on command, while leaving the operator's hands free. If you're using the pedal version,
insert the plug of the pedal power cord into the appropriate socket located next to the equipment, push the second switch on the position II, so start
and stop the suction operation by pressing relasind the pedal.
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the white ring fitted on the tank holder
for a more comfortable insertion of the same container.
Connect the short tube with the throttle connector marked as "VACUUM" fitted on the lid of the disposable pocket. The container should be positioned
on ON. The long silicon tube should be connected with the connector marked as "PATIENT".
In order to perform suction, close the connector marked as "TANDEM" with the lid provided.
If using the equipment with FLOVAC® disposable collection system antibacterial filters are not required since each disposable pocket is already
provided with an inside filter. The hydrophobic antireflux and antibacterial filter provided will protect the equipment, while also acting as float valve
deactivating vacuum generation when the maximum volume level is reached.
When the float valve intervenes signalling the device is too full, the suction source must be disconnected within no more than 5 minutes. Warning:
The accidental inversion of connections may cause contamination for the operator and/or for the vacuum generation equipment.
A wrong connection causes immediate destruction in case of contact with sucked liquids.
IN
INSTRUCTIONS
FLOW DIRECTION
IN / Fluid Inside
16
Jar Air Tube