METASYS AMALGAM SEPARATOR
2
1
58
3
RESET
CONTROL PANEL
1.
GREEN LIGHT:
Power pilot lamp
Normal condition:
When the equipment is connected, the green pilot lamp 1 lights
continuously.
Fault:
After the equipment connection, if the green pilot light 1 does
not light, check fuse and plate and, if necessary, replace them.
2.
ORANGE LIGHT:
Centrifuge control
Normal condition:
As a rule, this pilot lamp should not light. In case it lights, turn
the dental unit on and off several times (max. 5 times), in order
to check whether any fault occurred. If light does not extinguish,
a fault is present.
Fault:
A system controls power input to centrifuge's motor. When
input is too high, the light 2 starts to operate, to indicate the
presence of a faulty condition. At the same time, the suction
valve does not open and none operation can be carried out.
-
That condition can take place even when filter is removed
from its box, because some large particles could have
clogged and blocked the centrifuge.
-
The same condition can take place when a motor's fault is
present.
-
In order to eliminate any fault, replace the centrifuge set.
3.
YELLOW LIGHT:
Control of amalgam level
Normal condition:
When the equipment is switched on, an electronic system
controls several activities, among which the amalgam level.
When a 100% level is obtained, the yellow pilot lamp lights
continuously and the warning signal can not be cut out any
more. At the same time, the suction solenoid valve does not
open and none operation can be carried out. Thus, it will be
necessary to replace the amalgam container, in order to allow
the equipment to operate properly.