Telecamera Module; Acta Light Cure Module - eurodent Iso Plus Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

TELECAMERA MODULE

ACTA LIGHT CURE MODULE

48
Lifting the instrument the light on the handpiece is switched on and the
camera begins to telecast live. Pressing once the foot-control lever
rightwards, the still is got at full screen. For coming back to the live
image mode, the lever of the foot-control should be pressed again.
For showing up to 4 images in the same page, the lever should be
pressed for about 3 seconds. The images are stored and visualised
together on the monitor. For coming back to one image mode keep
the foot-control pressed for more than 3 seconds.
This module has no programmable functions and is independent from
the foot control.
To start operation, press the button located on the instrument itself.
The operation time for each start is fixed at 30 s; an acoustic signal
gets activated every 10 s.
Press the button again to stop operation.
The fibre optics tip is autoclavable according to the indications
given at page 50.
1 2
CHECKING THE ACTA LIGHT CURE
Switch the instrument on and point it at the light sensor (No 1): the
green light (No 2) must lit on.
If the red light lits on instead, clean the filter and the fibre optic rod.
Replacement of the curing light bulb
1) Loosen the fastening screw.
2) Remove the front casing, by squeezing it slightly.
When the camera is put back, nothing appears on the screen;
otherwise if it was stored one or more images, this/these will
remain on the screen.
The handpiece cannot be autoclaved and cannot be immersed
in a liquid. It is recommended to use the appropriate hygienic
protections.
The position of the module cannot be changed.
Do not point the instrument light towards the eyes.
MODULES
MODULES

Hide quick links:

Tabla de contenido
loading

Tabla de contenido