ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The JC-1321 [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the JC-1321 [KD-558] should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The JC-1321 [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the JC-1321 [KD-558] should assure that it is used in such an
environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment - guidance
The JC-1321 [KD-558] uses RF energy only for
its internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The JC-1321 [KD-558] is suitable for use in all
establishments other than domestic and those
Not applicable
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Not applicable
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
78
Electromagnetic
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.