Range 1, Range 2, Range 3
(RANGE 1, RANGE 2, RANGE 3):
Here you can enter the limits for each
of the weighing ranges. When a limit is
exceeded, the accuracy changes.
The following applies when entering
range limits:
Range 1 < range 2 < range 3 < max.
cap.
Thus the weighing capacity can be
divided into 4 ranges. The display reso-
lution changes in intervals of 1, 2, 5,
10, 20 etc. The lowest resolution is the
smallest scale interval d. Ranges you do
not require should be set to 0.
A/D Converter Configuration Menu
The following menu items for configuring the A/D converter are appended to the COMSPEC numeric menu when the service mode is active:
ADU-CON
STAND.
RANGES
8
Available Weight Units (Menu Item:
UNITS)
The weight units available for use in
the normal weighing mode are selected
here. Available units are indicated by
a circle ( o ) on the display (more than
one can be selected).
When using the instrument in legal
metrology, make sure the selected unit
is permissible.
Saving Configuration Parameters
(Menu Item: SAVE)
Once you have selected all required
parameters, select the SAVE menu item
to store the settings.
SINGLE
WEIGHTS
D
MAX
MULT.Int.
WEIGHTS
D
RANGE 1
RANGE 2
RANGE 3
MAX
MULT.-R.
WEIGHTS
D
RANGE 1
RANGE 2
RANGE 3
MAX
Checking and Configuring the Equip-
ment for Use in Legal Metrology
A metrological ID tag is included with
the display and control unit.
Following ADC configuration, write
the metrological data on this tag for all
weighing ranges. Affix this tag below
the display and cover it with the water-
proof acetate overlay.
Signum2
Standby
Scale
Check the settings under menu item 1.7
to make sure only the permitted weight
units are accessible.
A/D converter configuration
Standard configuration
Selection of weighing ranges
Single-range mode
Scale interval
Enter a value
Maximum capacity
Enter a value
Multi-interval mode
Scale interval
Enter a value
Range 1 limit
Enter a value
Range 2 limit
Enter a value
Range 3 limit
Enter a value
Maximum capacity
Enter a value
Multiple range mode
Scale interval
Enter a value
Range 1 limit
Enter a value
Range 2 limit
Enter a value
Range 3 limit
Enter a value
Maximum capacity
Enter a value
CF
REF
On
OK
Info
Clear
Toggle
Reference
Function
0
T
Fn
ISO-
n
Test
Tare
Zero
Function
Print
A
x10
Resolutio