Amprobe digital Ω Instrucciones De Uso página 4

Voltímetro bipolar con indicación lcd
Idiomas disponibles
  • ES

Idiomas disponibles

English
Thank you for purchasing your AMPROBE
detect DC and AC voltages of between 12 and 690V, for polarity, phase sequence and throughput testing at up to 2 k!, and for
FI/RCD tests.
®
The AMPROBE
digital! carries a high protection rating (IP65), making it suitable for use in harsh conditions.
Notes on safety:
You have selected a device that is designed to offer you a high degree of safety. It conforms to standards DIN VDE 0682-401
and IEC/EN 61243-3. In order to ensure that the unit is operated correctly and safely, read these operating instructions
thoroughly BEFORE operating the device.
The following safety precautions should be observed:
Check the voltage tester for correct functioning immediately before each use (as per VDE specification 0105, part 1),
by connecting it to a known voltage source (e.g. a 230V power socket). If the indicator shows that one or more
functions are not working correctly, do not use the device any further, but have it examined by a qualified service
technician.
Always hold the device by its handles. Avoid touching the probe tips.
Note that all testing for no voltage should be carried out at two poles.
Trouble-free operation is guaranteed at temperatures of between -10°C and +50°C.
Keep the device clean and dry at all times. The casing may be cleaned by wiping with a damp cloth.
General points:
Voltages have priority. If no voltage is detected at the probe tips (< 3.0V), the device is in throughput testing mode.
1.) Function:
The device is activated and shut down at the "ON/OFF" switch. (See drawing) The "Auto Power Off" function of the device
activates after about seven minutes.
2.) Self testing:
Hold the probe tips together when carrying out this test. The buzzer on the test device should sound clearly and the display
should show "000". If the LCD screen is blank or dimly illuminated, the batteries must be replaced.
3.) DC voltage testing:
When the probe tips are connected to a DC voltage within the nominal voltage range, the reading is displayed in volts and the
"DCV" LED indicator lights up. If a negative voltage is detected at probe tip "L1", a "-" (minus) symbol is displayed before
the reading.
4.) AC voltage testing:
When the probe tips are connected to an AC voltage within the nominal voltage range, the reading is displayed in volts and the
"ACV" LED indicator lights up.
5.) "Data Hold" function:
Press the "Data Hold" button to save the LCD screen reading to memory. The "Data Hold" function, which is indicated by a
red LED in the display field, can be toggled by repeated pressing of the same button.
6.) Phase testing:
Bring one of the probe tips into contact with a conductor, while touching the finger contact located at the back. When a phase
of min. 100V~ is connected, the small LCD screen indicates "L1".
7.) Rotating field testing:
Identify a phase conducting wire as described in point 6. Now connect two phase conductors to the probe tips and touch the
finger contact. If the phase connected to probe tip L2 follows the phase connected to probe tip L1, the phase sequence is right-
handed. The small display indicates "L1". If this is not the case, there is a left-handed field. The voltage indicator should give a
reading of approx. 400V. If "L1"appears in the small LCD screen and the only reading is 230V, only one phase conductor is
connected.
Operating Instructions
®
digital! two-pole voltage tester with LCD screen. The device can be used to
- 4 -
®
AMPROBE
digital!
loading