sPeCiFiCATiOns

valid at 23 deg. C after a 30 minute warm-up period
DC voltages
Measurement ranges:
500 mV, 5 V, 50 V, 500 V, 600 V
Resolution:
10 µV, 100 µV, 1 mV, 10 mV, 100 mV
Accuracy:
5 V, 500 V, 600 V:
±(0.05% of rdg.
500 mV, 50 V:
± (0.05% of rdg.
Overload protection:
V/ /T°/dB/
against COM and to chassis:
850 V
Low against chassis: 50 V
Input resistance:
50 V, 500 V, 600 V:
10 M II 90 pF
500 mV, 5 V:
>1 G II 90 pF
Input current:
10 A
CMRR
3)
:
100 dB (50/60 Hz ± 0.5%)
SMRR
4)
 60 dB (50/60 Hz ± 0.5%)
dB Mode
Accuracy:
±(0,0 dB +  digits)
(display > –8,7 dBm)
Resolution:
0.01 dB above 18% of fs.
DC current
Measurement ranges:
500 µA, 5 mA, 50 mA, 500 mA, 10 A
Resolution:
10 nA, 100 nA, 1 µA, 10 µA, 1 mA
Accuracy:
0,5 – 500mA:
±(0.% of rdg.
10 A:
±(0.% of rdg.
Voltage drop:
10 A range:
0. V max.
500 mA range:
.5 V max.
other ranges:
0.7 V max.
AC voltage
Measurement ranges:
500 mV, 5 V, 50 V, 500 V, 600 V
Resolution:
10µV, 100 µV, 1 mV, 10 mV, 100 mV
Accuracy 0,5-50 V:
0 Hz - 5 kHz:
± (0.% of rdg.
0 Hz - 0 kHz:
± (1% of rdg.
500 V and 600 V:
0 Hz - 1 kHz:
± (0.% of rdg.
0 Hz - 1 kHz:
± (1% of rdg.
Overload protection:
V/ /T°/dB/
against COM and to chassis:
850 V
Low against chassis: 50 V
Input impedance
AC mode:
AC + DC mode: 10 M II 90pF
Bandwidth at -3 dB:
80 kHz typical
dB Mode:
0 Hz - 0 kHz
Accuracy (–23.8 - 59.8 dBm): ±0. dBm
Resolution:
0.01 dB above 9 mV
CMRR
3)
:
 60 dB (50/60 Hz ± 0.5%)
Crest factor:
7 max.
AC current
Measurement ranges:
500 µA, 5 mA, 50 mA, 500 mA, 10 A
Resolution:
10 nA, 100 nA, 1 µA, 10 µA, 1 mA
1)
+0.00% of fs.
)
)
1)
+ 0.00% of fs.
)
)
at max.60 Hz or 600 V
p
DC
at max. 60 Hz or 50 V
rms
DC
)
1)
+ 0.00% of fs.
)
),
1)
+ 0.00% of fs.
)
)
1)
+ 0.07% of fs.
)
)
1)
+ 0.07% of fs.
)
)
1)
+ 0.07% of fs.
)
)
1)
+ 0.07% of fs.
.)
)
at max.60 Hz or 600 V
s
DC
at max. 60 Hz or 50 V
eff
DC
1 M II 90pF
S p e c i f i c a t i o n s
Accuracy:
0.5-500 mA: ±(0.7% of rdg.
10A:
± (1% of rdg.
1)
AC + DC measurements
Same as AC + 5 Digits
Resistance
Ranges:
500 , 5 k , 50 k , 500 k , 5 M , 50 M
Resolution:
10 m , 100 m , 1 , 10 , 100 , 1 k
Accuracy:
500 to 500 k :
±(0.05% of rdg.
5 M and 50 M :
±(0.% of rdg.
Input protected to max. 00 V
Measurement current
500 – 5 k -range:
1 mA
50 k -range:
100 µA
500 k -range:
10 µA
5 - 50 M -range:
100 nA
Measurement voltage: 10 V typical for open inputs, depending
on the value of resistance to be measured. Negative polarity of
measurement voltage is across common terminal.
Temperature
-wire resistance measurement with linearization for
PT100 sensors as per standard EN60751
Range:
–00 °C to +500 °C
Resolution:
0.1°C
Measurement current: ca. 1 mA
Display:
in °C, °F
Accuracy:
–00 °C to +00 °C
± 0. °C + 0,0005 x T
+00 °C to +500 °C
± 0.5 °C + 0,0005 x T
(T in °C, sensor tolerance not included)
Temperature coefficient
V =
500 mV, 50 V
600 V range
other ranges
V ~
600 V range
other ranges
mA
all ranges
mA-
all ranges
5 M , 50 M -ranges
other ranges
Miscellaneous
Operating temperature: +5°C ... +0°C
Storage temperature:
–0°C ... +70°C
Max. relative humidity:
5% ... 80% (without condensation)
Power supply (from HM8001-2):
+ 5 V
00 mA
~6 V
10 mA
Size (W x H x D):
15 x 68 x 8 mm
(without flat -pole connector)
Weight:
ca. 0.5 kg
1)
rdg. = reading;
)
= Common mode rejection ratio;
)
fs. = full scale;
Included in delivery: HM801 Multimeter, Manual,
HZ1 Interface cable, HZ15 PVC Test Leads, Software-
CD. Accessories: HZ10 Silicone Test Leads, HZ81
PT100 Temperature Sensor
Subject to change without notice
1)
+0.07% of fs.
)
) for 0Hz-5kHz
+ 0.07% of fs.
)
)
1)
+0.00% of fs.
)
+50m )
1)
)
+0.00% of fs.
)
eff
0 ppm/°C
80 ppm/°C
0 ppm/°C
80 ppm/°C
50 ppm/°C
00 ppm/°C
00 ppm/°C
00 ppm/°C
50 ppm/°C
1