Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Emission test
RF emissions
CISPR11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions IEC
61000-3-3
Guidance and manufacturer declaration - Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD) acc. to
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Comment: U T is the a.c. supply voltage prior to application of the test level.
Compliance
Electromagnetic environment – Guidelines
Group 1
The device uses RF energy only for its internal
function. Therefore, RF-emission is very low and it
is unlikely that any interference in nearby electronic
equipment.
Class B
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Class A
Symmetrical three-phase devices and other devices.
Passed
IEC 60601
test level
± 6 kV
contact discharge
± 8 kV
air discharge
± 2 kV for mains
cables
± 1 kV for input
and output lines
± 1 kV voltage
phase – phase,
± 2 kV voltage
phase – earth
< 5% U T ,
(> 95% dip in U T )
for 1/2 period
40% U T ,
(60% dip in U T )
for 5 periods
70% U T ,
(30% dip in U T )
for 25 periods
<5% U T ,
(> 95% dip in U T )
for 5 seconds
3 A/m
med 112655
Compliance
level
± 6 kV
contact discharge
± 8 kV
air discharge
± 2 kV for mains
cables
± 1 kV for input
and output lines
± 1 kV voltage
phase – phase
± 2 kV voltage
phase – earth
< 5% U T ,
(> 95% dip in U T )
for 1/2 period
40% U T ,
(60% dip in U T )
for 5 periods
70% U T ,
(30% dip in U T )
for 25 periods
<5% U T ,
(> 95% dip in U T )
for 5 seconds
10 A/m
2015-12-10
Electromagnetic environ-
ment – Guideline
Floors should be wood,
concrete or covered with
ceramic tiles. If floors are
covered with synthetic
material, the relative
humidity should be at least
30 %.
The supply voltage quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the device
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered by a
UPS (uninterruptible power
supply) or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.