If a sampling head with a bayonet connector is
t
used: Mount bayonet to Tri-Clamp (TC 50 .5)
adapter on the bayonet connection .
Insert a gasket for the Tri-Clamp connection in the
t
1
groove (1) of the connector piece .
NOTICE Damage to the sensor when connecting
t
pipes or adapters! Insert the piping and adapters
only so far as to not allow the stainless steel
strainer and the sensor to be touched .
Connect the hose or piping to the air inlet of the
t
sampling head . Use suitable accessories, e .g .
connectors, adapters or Tri-Clamps (see
Chapter "6 .1 Accessories," page 20) .
Secure the connection using a Tri-Clamp .
t
4.5

Conducting Exhaust Air Away

The exhaust air that flows out of the sampling head air
outlet can be left in the ambient air or conducted into
an exhaust air system .
If the sampling head is operated in controlled areas
of clean room class A-B: The exhaust air should be
removed .
When the sampling head is operated in controlled
areas of clean room class C-D: It is recommended
that the exhaust air be removed if the particle
concentration in the exhaust air exceeds the
permissible air purity limit . To this end, observe the
applicable regulations on air purity, e .g . the valid
United States Pharmacopeia (USP), Eu .Ph or ISO
guidelines .
MD8 Airscan Sampling Head Installation Instructions