Filter Adapter; Airborne Microbe Collection - Sartorius MD8 Airscan 16746-01-SHB Serie Instrucciones De Instalación

Unidad de aspiración para colector de gérmenes aéreos
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 79
3.7

Filter Adapter

The filter adapter accommodates the filter and, depending on the model, is suitable
for accommodating gelatine membrane filters or cellulose nitrate membrane filters .
Fig . 6: Filter adapter for gelatine membrane filters
No.
Designation
1
Filter mount
2
Attachment for bayonet
The filter adapter must be placed on a bayonet connection . An adapter with bayonet
connection is additionally required for sampling heads with a Tri-Clamp connection
(TC 50 .5) .
3.8

Airborne Microbe Collection

Pumping of Air Sample Volume (Isokinetic Sampling)
The sampling head generates a vacuum and pumps a defined air sample volume from
the ambient air . The air flow can be adapted to the ambient speeds so that the flow
velocities in critical areas, e .g . in clean rooms, are not disturbed and turbulences are
avoided .
Retention of Biological Particles
Active airborne microbe collection takes place in the filter . The biological particles
from the air, e .g . viruses, bacteria, yeasts and fungi, are retained in the filter . This
retention is almost independent of the configured air flow .
Description
Accommodates the gelatine membrane filter .
Is placed on the bayonet connection .
MD8 Airscan Sampling Head Installation Instructions
Device Description
1
2
11
Tabla de contenido
loading

Este manual también es adecuado para:

Md8 airscan 16746-01-sht serie

Tabla de contenido