Functional Elements; Operation; Time Programmes; Rotary Plate - Dreve Dentamid EyeVolution MAX Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

EyeVolution
®

6. Functional elements

8
6
5
No. Description
1
Power supply connector socket
2
Main switch
3
Display foil
4
Time programme push-button
5
Rotary plate function push-button
6
Function display
7

Rotary plate

8
Reflective foil
9
LED circuit board
10
Glare protection
11
Cover handle
12
Device cover
13
Inspection window
22
MAX
13
12
11
10
9
7
3
4
2
1

7. Operation

Activate the main switch (2) on the back
lum plate is recommended for exclusively
of the device; the functional display LED
occlusal modelling (e.g. inlay or onlay).
(6) lights up in red. The device is now
ready for operation.
Open the polymerisation chamber by
raising the device cover (12) at the han-
dle (11). The opening angle is approx.
55° and is fixed in its final position by
a magnet.
The central point of the surface that is to
be polymerised must always be positioned
on the seating for the model at a distance
of 2-3 cm from the middle LED that has a
Release the cover from its magnetic
black circle round it.
anchoring by pulling slightly on the han-
Fixing the model vertically on the pendu-
dle and close the polymerisation chamber.

7.1 Time programmes

The EyeVolution
MAX light-curing
Opaquer: 1:00–2:00 min
®
device features several pre-programmed
Pre-polymerisation: 0:30–1:00 min
time relays. Use the push-button (4) to
Interim polymerisation: 1:00–2:00 min
activate time programmes with 1 / 2 / 3 / 5
End polymerisation: 2:00–5:00 min
and 10 minutes.
Start the selected time programme by
The following times apply, depending
pressing the corresponding push-button
on the composite material that is being
(4). The polymerisation process starts
used and for a maximum layer thickness
automatically and is indicated as an active
of 2 mm:
programme by a green LED.
7.2 Rotary plate
and is automatically activated when the
A reflective pendulum plate (7) was
polymerisation programme starts. The LED
designed for the seating of the model to
(6) indicates the operating status of the
ensure optimum polymerisation with as little
rotary plate:
shadowing as possible. Starting in a central
Green LED = rotary plate in action
position (see point 7.1), the seating of the
Red LED = rotary function deactivated
model first moves to the right by 90° after
Use the push-button (5) to activate or deac-
the program is started and then shifts into a
tivate the rotary plate function at any time,
180° pendular motion. The oscillation func-
even during on-going polymerisation pro-
tion is coupled with the time programmes
grammes.
optimal
start position
23