• For post-curing of the parts, please select the parameter set-
tings for your process. You can find more qualified printers on
our website www.dreve.dentamid.de.
• Post curing
PCU LED N
2
Rapid Shape D,DII / Asiga MAX UV
6 minutes, 80 %
light power (vacuum)
FotoDent® flash
Rapid Shape D, DII
8 minutes + 1 minute gas pre-flow
Asiga MAX UV
10 minutes + 1 minute gas pre-flow
RS Cure
Rapid Shape D, DII / Asiga MAX UV
6 minutes, 75 %
light power (vacuum)
• Impurity due to the construction or a break of the material and
impurity due to operation mistakes cannot be excluded. Thanks to
the low viscosity however it is possible to filtrate FotoDent® cast.
It is recommended to regularly take out the container of the
production unit, to homogenise and filtrate the content.
Upcoming entrained bubbles get eliminated through a holding
time of approx. 1–2 hour with FotoDent® cast.
When embedding FotoDent® cast, please note the manufacturer´s
data of the used embedding materials. To ensure that the resin
combusts without any residue, a holding time of 60 minutes at
800–900 °C is recommended.
To avoid damage of material quality do not expose the liquid
material to light under any circumstances. Variations from the
described manufacturing process may lead to modified mechan-
ical characteristics and/or colour variations of the FotoDent® cast
material.
Possible dangers
Please note the safety instructions for processing FotoDent® cast.
Hazard statements: H317 May cause an allergic skin reaction.
Precautionary statements: P261 Avoid breathing dust/fume/gas/
mist/vapours/spray. P280 Wear protective glovesP304+P340 IF
INHALED: Remove person to fresh air and keep comfortable for
breathing. P305+P351+P338 IF IN EYES: Rinse cautiously with water
for several minutes. Remove contact lenses, if present and easy to do.
Continue rinsing. P501 Dispose of contents/container in accordance
with local/regional/national/international regulations. P312 Call a
POISON CENTER/doctor if you feel unwell
Contains: Diurethandimethacrylate, triethylene glycol dimeth-
acrylate, phenyl bis(2,4,6-trimethylbenzoyl)-phosphineoxide
Lot number / Expiry date
The lot number and the expiry date are indicated on each
FotoDent® cast packaging. In case of claims please always indicate
the lot number of the product. Do not use the product after the
expiry date.
Lieferform / Form of delivery
D35100
FotoDent® cast
1,0 kg Flasche/bottle
Rx only
Shake
Achtung
before use
Warning
Stand der Informationen / Date of information: 03.2019
Dreve Dentamid GmbH
Max-Planck-Str. 31 • 59423 Unna/Germany
FotoDent® cast
385 nm / 405 nm
Gebrauchsanweisung / Directions for use