omegaSoft Software
systeM reQuireMents
OmegaSoft software is designed to run under Windows. Suitable versions are
Windows 2000, Windows XP, Windows Vista and Windows 7.
The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
installation
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
note Do not connect the os150-usb sensor to the pc before the software is
installed
connectinG the os150-usb sensor to a pc
1. Connect the circular connector on the USB cable provided to the OS150-USB sensor
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the OmegaSoft desktop icon
note Do not disconnect the usb cable before closing down the omegasoft
program

configuration

Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
Setting Output Processing
The temperature range for the 4 to 20mA output, the averaging period and the peak/
valley hold processing can be set by using the main program menu to select:
[Setup] → [Output Processing]
The temperatures corresponding to 4 and 20mA can then be set (minimum –40°C,
maximum 1000°C, minimum span 100°C, maximum span 1000°C).
To minimise the the effects of temperature fluctuations, noise etc on the output
signal, the averaging period can be set between 0 and 60 seconds.
If required, Hold Processing can be applied by setting hold mode to Peak Hold or
Valley Hold and setting a hold period between 0 and 1200 seconds.
Setting Target Emissivity
To set an Emissivity value for the OS150-USB Sensor, use the main program menu
to select:
[Setup] → [Emissivity]
Emissivity can then be set to a value between 0.1 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
8