Health O Meter 600KL Instrucciones Para El Usuario página 13

Ocultar thumbs Ver también para 600KL:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

OPERATING INSTRUCTIONS (CONT.)
Using the Scale for Height Measurement Only
1. Press the ON/OFF button to turn the scale on.
2. Wait until '0.0' and 'Zero' appear on the left side of the display.
3. Press the WEIGHT/HEIGHT/BMI button once. The scale will display the position of the height
rod.
4. Ask the patient to step on the scale platform. Adjust the height rod so that the headpiece gently
rests on the patient's head. Read the height.
5. Press the RESET button to revert back to weight display.
To Obtain BMI
1. Press the ON/OFF button to power on the scale.
2. Ask the patient to step on the scale platform and obtain the patient's weight. Press the
WEIGHT/HEIGHT/BMI button once. The scale will now display the position of the height rod.
While the patient remains on the platform, adjust the height rod so that the headpiece gently rests
on the patient's head. Press the WEIGHT/HEIGHT/BMI button again. The display will show the
patient's BMI.
3. If desired, press the SEND button to send the weight, height and BMI information to the scale
USB port.
4. Press the RESET button to revert back to weight display.
Locking the Weight Measuring Unit in KG (600KL only)
Caution: Please make sure you carefully and precisely follow the directions below as you will
be in the PROGRAM MODE of your scale. Any modification outside of what is listed below
may greatly affect the functionality of your scale.
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNITS button twice, and then the ZERO button once again. The scale will
display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-0 or OP1-1 will be displayed.
4. Press the
once and the display will read "OP2-0" for the KG lock function, then press the
UNITS button to enable KG lock. Your display should read "OP2-1".
until display reads "DONE"
5. Press the
6. Press the SEND button, the scale will read "START", then read "0.0 Kg"
You have successfully locked your scale into KG Only.
Deactivating KG Lock (600KL only)
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNITS button twice, and then the ZERO button once again. The scale will
display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-0 or OP1-1 will be displayed.
once and the display will read "OP2-1"for the KG lock function, then press the
4. Press the
UNITS button to disable KG lock. Your display should read "OP2-0".
5. Press the
until display reads "DONE"
6. Press the SEND button, the scale will read "START", then read "0.0"
You have successfully deactivated the KG lock on your scale.
P/N UM600KL_600KG_Rev20190416_Web