VisualSonics Vevo MD Manual De Instrucciones página 28

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

START
Non-critical cleaning and disinfecting
1 Put on gloves, for protection from cleaning and disinfecting materials. Change
gloves as needed throughout the procedure.
2 Unplug the foot switch from the system.
3 Obtain a compatible cleaner. Compatible cleaning materials are listed in the table
Compatible cleaners and disinfectants
Caution:
Make sure the solution does not get into any of the connector ports.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces if not using a wipe.
4 Clean the foot switch and cable to remove any visible gel, debris or bodily fluids.
♦ Wipe from clean areas to the soiled areas to avoid cross-contamination.
WARNING:
Do not wipe the USB connector on the foot switch cable.
5 Verify that all gel, debris and bodily fluids have been removed from the foot switch
and cable. If necessary, repeat Steps 4, with a new wipe.
6 Obtain a compatible disinfectant. Compatible disinfectant materials listed in the
table
Compatible cleaners and disinfectants
Caution:
Make sure that the solution does not get into any of the foot switches.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces, if not using a wipe.
28
Cleaning and Disinfecting - English
on page 26.
on page 26.
Tabla de contenido
loading

Tabla de contenido