Citizen CHU305 Manual De Instrucciones página 24

Ocultar thumbs Ver también para CHU305:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Appendix
Guidance and manufacturer's declaration - electromagnetic emissions
The [CHU305] is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the [CHU305] should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/ fl icker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The [CHU305] is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the [CHU305] should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic fi eld
IEC 61000-4-8
Note: U
00_CHU305_USA.indb Eng-24
00_CHU305_USA.indb Eng-24
ELECTOROMAGNETIC COMPATIBILITY INFORMATION
The use of accessories and cables other than those specifi ed (other than CITIZEN
original parts) may result in increased emissions or decreased immunity of the unit.
Compliance
Group 1
Class B
N/A
N/A
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±1 kV
line to line
±2 kV
line to earth
<5 % U
T
(>95% dip in
U
) for 0.5 cycle
T
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycle
<5% U
T
(95% dip in U
)
T
for 5 sec.
3 A/m
is the A.C. mains voltage prior to application of the test level.
T
Electromagnetic environment - guidance
The [CHU305] uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The [CHU305] is suitable for use in
all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance
Electromagnetic environment -
level
guidance
±6 kV contact
Floors should be wood, concrete or
ceramic tile. If fl oors are covered with
±8 kV air
synthetic material, the relative humidity
should be at least 30%.
N/A
N/A
N/A
N/A
N/A
N/A
Power frequency magnetic fi elds should
3 A/m
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
- Eng 24 -
2015/01/13 19:53:22
2015/01/13 19:53:22
Tabla de contenido
loading

Tabla de contenido